Kyoritsu Reagent Set for Water Analyzer Silica (Low Range)

SKU
LR-SiO2D
฿0.00
Pre-Order 30-45 days
Out of stock

To measure Silica ion (SiO32-) concentration in the sample.
Measureing Method: Molybdenum Blue Method
Range: 0.30 ∼ 5.00 mg/L
Reaction Time 10 min.
Kyoritsu-Chemical, Japan

Kyoritsu Reagent Set for Water Analyzer Silica (Low Range) (SiO32-)

Product Name Reagent Set for Water Analyzer No.20D Silica (Low Range)
Model LR-SiO2D
Color Change None → Light blue → Blue
Measureing Method Molybdenum Blue Method
Range 0.30 ∼ 5.00 mg/L
Reaction Time 10 min.
Quantity (pcs./box) 30
Outer Box Size 160L × 35W × 107H mm
Total Weight approx. 75 g

 

1 - 2 3 4 5 6
  • Press SiO2-D
  • Press <Enter> to change over to the display of measurement procedure.
  • Fill the cell with 25 mL of sample (up to the line).
  • Insert the cell into the cell box with ▼ of cell facing ▲ of cell box. Press <Blank>.
  • Take out the cell from the cell box and add 1 mL of R-1 Reagent with the attached dropper into the cell.
  • Cap the cell tightly and shake the cell 5 - 6 times. Wait for 4 minutes.
  • Add R-2 reagent into the cell.
  • Cap the cell tightly and shake the cell 5 - 6 times. Wait for 4 minute.
  • ใAdd R-2 reagent into the cell and press Start.
  • Cap the cell tightly and shake the cell 5 - 6 times.
  • Before 5 minutes pass, insert the cell into the cell box immediately with ▼ of cell facing▲ of cell box.
  • After 5 minutes passed, the measurement will be displayed automatically.
SKULR-SiO2D
Weight0.070000
Dimension160L × 35W × 107H mm
Manufacturerkyoritsu-chemical