Signatrol SL7000 Series Universal Input Data Loggers

SKU
SL7000
Out of stock

Universal Input Data Loggers
PT100, Thermocouples, mA, mV, Volts, Slidewire
8 LED's indicator status for logging & alarm
Up to 128,000 Readings
Up to 6 Years Battery Life

Signatrol SL7000 Series Universal Input Data Loggers

How to Order SL7000 (Order Codes)

Customer can be select specification below and input number into order codes.

  SL7              0           
Series 0. No Display
1. Display
0. USB Interface 0. Internal Temperature only
1. Internal Temperature + Internal RH
2. 2 Universal Inputs + Internal Temperature
3. 2 Universal Inputs + Internal Temperature + RH
4. 4 Universal Inputs + Internal Temperature
5. 4 Universal Inputs + Internal Temperature + RH

Measuring Ranges & Accuracy

Measurement Range Accuracy
Internal Temperature -30 to 70 °C ± 0.5 °C
Internal Humidity 0 to 100% RH ± 2.0% RH over 10 to 90%
± 4.0% RH other
PT100 (EN 60751) -200 to 700 °C (Wide Measuring Mode)
-200 to 260 °C (Narrow Measuring Mode)
± 0.2 °C + 0.1% rdg (Wide Measuring Mode)
± 0.1 °C + 0.1% rdg (Narrow Measuring Mode)
PT100 (JISC) -100 to 457 °C (Wide Measuring Mode)
-100 to 260 °C (Narrow Measuring Mode)
± 0.2 °C + 0.1% rdg (Wide Measuring Mode)
± 0.1 °C + 0.1% rdg (Narrow Measuring Mode)
Thermocouple Type K -200 to 1370 °C ± 0.1% FSD ± 0.5 °C
Thermocouple Type J -200 to 1200 °C ± 0.1% FSD ± 0.5 °C
Thermocouple Type T -200 to 400 °C ± 0.1% FSD ± 0.5 °C
Thermocouple Type S 100 to 1760 °C ± 0.2% FSD ± 0.5 °C
Thermocouple Type R 100 to 1760 °C ± 0.2% FSD ± 0.5 °C
Thermocouple Type B 700 to 1820 °C ± 0.2% FSD ± 0.5 °C
Thermocouple Type N -180 to 1300 °C ± 0.1% FSD ± 0.5 °C
Ni120 -80 to 230 °C ± 0.2 °C ± 0.1% rdg
Ni100 -60 to 180 °C ± 0.2 °C ± 0.1% rdg
CU10 -200 to 260 °C ± 3 °C ± 0.5% rdg
CU100 -100 to 260 °C ± 0.2 °C ± 0.1% rdg
Thermistor TH-001 -40 to 105 °C ± 0.5 °C
Thermistor TH-002 -30 to 90 °C ± 0.6 °C (0 to 70 °C), ± 1.2 °C (-30 to 90 °C)
Thermistor TH-003 -35 to 105 °C ± 0.5 °C
Thermistor TH-004 -40 to 105 °C ± 0.5 °C (0 to 70 °C), ± 0.8 °C (-40 to 105 °C)
Slidewire 0 to 100% ± 0.02% FSD
mV -110 to 110 mV ± 0.01% FSD
10 Volt 0 to 11 VDC ± 0.01% FSD
1 Volt 0 to 1.1 VDC ± 0.01% FSD
4/20 mA 0.1 to 21 mA ± 0.01% FSD (for Isolated Loops)
Pulse Counter Up to 65,535 pulses per sample period 20 ppm

SPECIFICATION

Memory 0 or 2 Universal Inputs - 62,000 Readings
4 Universal Inputs - 128,000 Readings
Sample Rates 1 Second to 24 hours for all channels
Logging Options Spot Reading or Maximum, Minimum, Average over a number of readings.
Channel Indicators 8 Green LED's to indicate logging status, changing to Red, to indicate alarm conditions.
Interface USB
Low Battery Warning Low Battery displayed when battery life has less than one month to go.
Power Requirement Internal Battery 3.7V AA Lithium (life 6 years at 10 minute sample rate)
Dimension / Weight 124 x 78 x 32mm / 165g
SKUSL7000
Weight165.000000
Dimension124 x 78 x 32mm
ManufacturerSignatrol
Measurement objectAmbient Temperature, Core Temperature, Surface Temperature, Temperature in Liquid
Sensor connector typeMiniture connector for thermocouple
Supported sensor TypeNo
Other FunctionNo
Power SupplyInternal Battery
Resolution0.1 °C
Temperature Sensor TypeRTD (Pt100, Pt1000), Thermistor, Thermocouple
Number of Channel3 Channels and more...
Water resistanceNo
Sensor Output InterfaceUSB
Temperature RangeDepending on probe used
Data loggerYes
AccuracyDepending on probe used
Your Rating