TOPCON UVR-300 UV Radiometer

SKU
UVR-300
Out of stock

UV Radiometer
Three different types of high-sensitivity detectors
UD-250: 220 - 300nm
UD-360: 310 - 400nm
UD-400: 360 - 490nm
**Detectors aren't include**
TOPCON, Japan

TOPCON UVR-300 UV Radiometer

The UVR-300 is a hady-type UV radiometer with excellent efficiency and expansive ability. Three different types of high-sensitivity detectors, each for differing wavelength ranges, are available to deal with specific usage.

Measuring bactericidal lamps, measurement of quantity of light of photochemical reaction, measurement of quantity of light of photoresist, and testing solarization are just a few areas where the UVR-300 stands out.

  • EXtra digit for detecting sensitivity : 0.1 - 280,000μW/cm2
  • Measurement mode are sectable by Keyboard
  • USB Interface
  • Correction factor (C.C.F. mode)
  • Deviation measurement (◢ mode)
  • Percentage measurement (% mode)
  • Measurement of integral irradiance (mJ/cm2 mode)
  • Photochemical reaction
    • Photoresist for manufacturing semiconductors
    • Photosensitvie materials for printing or plate making
    • Photo color fading
    • Evalution of solar battery properties
    • Testing for deterioration of products
  • Photoelectric reaction
    • Exposure for electrophotography
    • Printing for electrophotography
  • Biological reaction
    • Erythema, pigmentation
    • Treatment of facula, diagnosis of photo-hypersensitivity
    • Aid and control in raising livestock or fish
    • Suppressing turion in plants
    • Photosynthesis
  • Measurement and analysis of germicidal effects in food processing
System Dimension
Display Range 0.1 - 280,000μW/cm2
Auto/Manual 4-step range
Display 4-digit LCD Read
Linearity ±5% of rdg.: ±1 digit (Auto range)
Measurement Wavelength Range UD-250: 220 - 300 nm
UD-360: 310 - 400 nm
UD-400: 360 - 490 nm
Angule Incident Light Characteristics UD-250: 30° ±3%, 60° ±15%
UD-360 / 400: 30° ±3%, 60° ±10%
Temperature Characteristics Within ±3% (-10 to 4°C : against 23°C)
Humidity Characteristics Within ±3%
Analog Signal Output 0 to 3Vmax, 1mV / 1 digit
Interface USB (Virtual COM port)
Power Supply AA battery ×2
Operating Conditions Temperature: -10 to 40°C
Humidity: 85% R.H. or less (no condensation)
Dimensions Approx. 195mm × 70mm × 33mm (including detector)
Weight About 260g (including detector and batteries)
Detection Element Silicon Photodiode
Calibration Light Source UD-250: GL-15
UD-360: FL20S•BLB
UD-400: Standard Light Source A
SKUUVR-300
Weight260.000000
Dimension195 x 70 x 33mm
ManufacturerTOPCON
Data loggerNo
Solar Power RangeNo
UV measurementUVA, UVB, UVC
Switching Lx RangeNo
UV measurement rangeYes
Lux measurement range lx
Other FunctionNo
Power SupplyBattery AA x 2
Resolution0.1
Water resistanceNo
Sensor Output InterfaceUSB
AccuracyDepending on specifications table by Sensor
Your Rating

Accessories Include:

  • UVR-300 (main body) ×1
  • Cap ×1
  • Leather Case ×1
  • CD-Rom (USB Driver / Instruction Manual) ×1
  • Analog Output Plug ×1
  • AA Battery ×2

Detectors:

  • UD-250 (220 - 300nm)
  • UD-360 (310 - 400nm)
  • UD-400 (360 - 490nm)

Options:

  • AC adapter ZV-42
  • Extension Cable
    • 2m ZV-21
    • 5m ZV-22
    • 10m ZV-23
    • 20m ZV-24
    • 30m ZV-25